The phaco systems light-weight, reliable and non-skid footpedal treadle is made of
water resistant and highly durable construction with a rugged, 10 foot polyurethane cable.
With it's easily recognizable audible and vibrational knock to indicate footpedal stages,
it outperforms its more cumbersome and less reliable competition.
PHACOEMULSIFICATION SYSTEM - FOOTPEDAL CONTROLS
Products Foldable intraocular lens PMMA iols Viscoelastics Phacoemulsification system Intra-stromal inlays CTRs Sutures Instruments Sterilizers Economy Links Search
Eye4Vision Int'l.
®
* Surgeon variable via treadle position
† Audible / vibrational knock indicates entry into new position
Phacoemulsification system
|
Phacoemulsification system
|
Phaco System Standard Footpedal Phaco System Cold Phaco Footpedal
|
►
U/S PULSE
-x- IRR IRR+ASP IRR+ASP+U/S * IRR+ASP+U/S * REFLUX
|
U/S
-x- IRR IRR+ASP IRR+ASP+U/S * IRR+ASP+U/S * REFLUX
|
I/A
-x- IRR IRR+ASP* IRR+ASP* IRR+ASP* REFLUX
|
CAUTERY
-x- -x- CAUT CAUT CAUT REFLUX
|
TREADLE Position 0 1† 2† 3† 4 REFLUX
|
FUNCTION vs MODE
Treadle
Position
C
C
VITRECTOMY ( S ) Single Cut -x- IRR IRR+ASP* IRR+ASP*/CUT IRR+ASP*/CUT REFLUX
|
VITRECTOMY
-x- IRR IRR+ASP*/CUT IRR+ASP*/CUT IRR+ASP*/CUT REFLUX
|
Function versus Mode
Irrigation pinch valve is actuated, permitting the elevated irrigation solution
to safely maintain intraocular pressure.
Enables footpedal proportional aspiration rate via peristaltic pump, permitting
required vacuum for effective aspiration.
Enables footpedal proportional ultrasonic ( U/S ) Power to be delivered to the attached
ultrasonic handpiece with maximum U/S Power at Position 4.
Same as U/S except U/S Power is
pulsed at 10 pulses/sec.
The vitrector is actuated at selected cut rate whenever
treadle position is ≥ Position 2.
Actuates reverse peristaltic pump rotation (20cc/min.)
Electrically heat targeted tissue held by the cautery forceps at the set cautery power
♦
Light-weight, reliable and non-skid.
Description
U/S Continuous
P
U/S Pulsed
CUT
REFLUX / CAUTERY
C
IRR - Irrigation
ASP - Aspiration
Footpedal
Actual vacuum and requested aspiration rates displayed whenever treadle ≥ Position 2
with maximum aspiration rate at Position 4.
Peristaltic pump actuation is only enabled when selected maximum vacuum has
not been exceeded.
Any vacuum generated by the peristaltic pump is automatically and safely vented
whenever treadle position is returned ≤ Position 2.
♦
♦
♦
Requested U/S Power is displayed whenever treadle position ≥ Position 3.
Aspiration rate is fixed at the selected aspiration rate.
♦